Elite Thermal’s Ash Fusion Determinator, EATC17 is the high-temperature floor-mount model, which works similarly to the EATC16plus, but for samples which fuse at higher temperatures. A maximum temperature of 1700°C can be achieved using heating elements consisting of molybdenum disilicide. The results obtained are similar for both EATC17 and EATC16 plus analysers.
General Specifications | EATC17 |
---|---|
Ash Fusibility Determination | Automatic |
Fusion Points | IDT (Initial deformation Temperature), ST (Softening/Sphere Temperature),
HT (Hemisphere Temperature) & FT (Fluid/Flow Temperature) |
Test Method | ASTM D 1857; ASTM E953; BS ISO 540; BS ISO 21404, CEN/TS 15370-1; CEN/TR 15404:2010.
DIN 51730; ISO 540; ISO 21404 |
Capable to Analyse | Cube/Cylinder, Pyramids/Cone, Upright cone/Upright Pyramid and Truncated cone/Truncated Pyramid. |
Sample shape identification | Automatic - Cube/Cylinder, Pyramids/Cone, Upright cone/Upright Pyramid and Truncated cone/Truncated Pyramid. |
Analysis atmosphere | Oxidizing Atmosphere/Reducing atmosphere |
Furnace Specifications | |
Temperature range | up to 1700°C |
Temperature Ramp Rate | programmable 1°C to 12°C per minute |
Temperature precision | ±5°C as per standard test methods at 1064 C (99.98% pure gold wire sample melting point) |
Number of heating elements | 6 nos - High temperature resistance type heating elements |
Working tube dimensions | 86 x 76 x 675mm |
Material of construction of working tube | High grade RCA Alumina work tube |
Analysis Time | 4 hours typical cycle time (depending ramp rate and temperature range) |
Stand by Temperature | |
Stand by Temperature | Room Temperature |
Ventilation | |
Ventilation | Forced air ventilation |
Exhaust | Pipe to be vented into a separate fume hood |
CO Monitor (Optional on request) | Integrated CO monitor with auditory alarm, Gas flow shut off on alarm. This is factory installed option. Need to order along with main EATC17 Instrument. |
Gas requirements | |
Gas requirements | Integrated four gas flow meters to enable automatic switching of the gases based on the selected test conditions, such as oxidation or reduction. |
Electrical requirements | |
380 – 415 V, 50/60 Hz two phase 25 A | |
Environmental Conditions | |
Operating Condition | 15°C to 35°C |
Relative Humidity | 20% to 80%, non-condensing. |
PC specifications | |
Required PC Specifications | Processor: i3 or i5, RAM: Minimum 4GB, Memory: 512GB HDD or SSD PCI slots: Minimum 1, PCIe slots: Minimum 1, RS232 Ports: 2 OS: Windows XP or higher |