Bottom Loading Chamber Furnace

Automatic Plastometer PF-12

Overview

The PF-12 plastometer is intended for measuring plasticity properties of coal heated under the conditions defined by the ASTM D 2639 – 08 and ISO 10329 standard. The outcomes of the test are plasticity nature of the coal.

The PF-12 plastometer is a fully automatic, compact, single furnace instrument. A single furnace with molten solder bath is equipped with sophisticated control system of heating range and heating rate to ensure the required accuracy level to meet the requirements of ASTM D-2639-08, ISO 10329. For quick cooling, the furnace has a built-in fan (allowing another test to be started shortly afterwards).

Automatic Plastometer PF-12
  • Temperature (deg C) and Fluidity (D.D.P.M) and Stirrer speed (RPM) are displayed at the end of each test.
  • Automatic crucible movement (up & down).
  • Built – in exhaust device.
  • The temperature measurement failure detection.
  • Cooling of the furnace by the fan analysis.
  • Automatic creation of *. csv files for statistical analysis.
Temperature Range 200°C to 600°C (typical)
Heating Rate Programmable 1 deg C/min to 9 deg
C/min
Temperature Accuracy ±0.1°C for temperature
Thermocouple k-type thermocouple
Torque Measurement Range 101.6 ± 5.1 g.cm (40±2.2 g.inch)
Measured Plasticity Range 30000 DDPM
Plasticity Measuring Resolution 0.1 DDPM
Power Rating 230V±15%, 50/60 Hz, 1050W