Volatile Matter Furnace, VMF/ISO
"

Volatile Matter Furnace - VMF/ISO

Overview

This is a chamber furnace that is designed for Volatile Matter Analysis as per ISO 562.This furnace offers temperature and response times to perform Volatile matter determination.

Volatile Matter Furnace, VMF/ISO
  • Maximum Temperature : 1100°C
  • Continuous Temperature : 1050°C
  • Chamber (mm) - H x W x D -110 x 200 x 254
  • Heating elements (Embedded in ceramic slabs) located on 4 sides of the chamber ensure rapid heating required for analysis as per ISO562.
  • The chimney is fitted with a valve to restrict airflow/convection through the furnace.
  • Provision for inserting external thermocouples (Three positions) to check the temperature under the crucibles is provided.
  • Protection of the elements from carbon build-up or corrosive atmosphere, inherent in the slab design makes it ideal for volatile matter analysis.
  • Vertical lifting door keeps the hot face away from the operator when the door is opened
  • Positive break door safety switch isolates heating elements from power supply when door is opened
  • High-end micro processor PID controller.
  • Over temperature protection controller.
  • Multi segment, multi program storage controllers with audible alarm & timer.
  • 4,6 & 9 crucible rack with loading handle available.
  • Crucibles and lids available as per ISO 562..
  • Calibrated thermocouple probe with digital indicator.