Thermogravimetric Analyser et500

Thermo Gravimetric Analyser - TGA et500

Overview

Elite's TGA et500 is a Dual Carousel Thermogravimetric Analyser. The unique capability of the TGA et500 involves its control over crucible covers. This thermogravimetric analyser has the ability to both place and remove crucible lids during the analysis process. This functionality enables precise determination of volatile content in substances such as coal and coke, among other applications.

Thermogravimetric Analyser et500
  • Powerful high performance Thermogravimetric Analyser
  • Fully automated analysis is capable of processing up to 19 samples.
  • Capable to accept maximum sample weight up to 5 grams, higher range can be offered on request the heating ramp rates can be programmed according to the prescribed test method.
  • TGA et500 employs a dual carousel design for automatic placement and removal of crucible covers
  • Automatic Positioning and removing of Crucible Covers Operation of the analyser is controlled by a Windows-based application.
  • The TGA et500 instruments are integrated with top loading balance with inbuilt auto calibration facility and the ability to weigh the test specimens and crucibles repeatedly throughout the analysis
  • An integrated balance with a readability of 0.1 mg allows mass determination with high accuracy and reliability.
  • Cooldown process is initiated subsequent to each Analysis Cycle.
Specifications TGA et500
Temperature Range Programmable from ambient to 1100°C
Temperature Control Precision ±2 deg C (or) ±2% of set point temperature
Temperature Stability ±2 deg C (or) ±2% of set point temperature
Ramp Rate Programmable from 10°C /minute to 500°C / minute
Balance Integrated Balance
Balance Resolution 0.0001g (0.1mg)
Balance Readability 0.0001g (0.1mg)
Weight Loss 0-100%
Sample Size Upto 10 grams based onthe sample type and its characteristics
Number of Samples 19 Samples +1 Reference
Number of Carousels Two (one for Crucibles, and the other one for Crucible lids)
Carousel Material Metal or Ceramics
Weighing Precision 0.02% RSD (on inert Samples)
Electrical Power Requirements 230V (± 10%) / single phase / 50/60Hz / 15A
Computer 230V (± 10%) / single phase / 50/60Hz / 2A
Item Name Part Number
Single Furnace Thermogravimetric Analyser with two carousels TGA et500
Double Furnace Thermogravimetric Analyser with four carousels TGA et500D
Ceramic crucibles et100-1460
Ceramic crucible covers et100-0460
Ceramic carousel for Crucible covers, 20 Positions et107-3210
Ceramic carousel for Crucibles, 20 positions et107-3730
Metal Carousel for Crucible covers, 20 positions et112-2540
Metal Carousel for Crucibles, 20 positions et112-2550